2D axial slice, 2.5 mm grid. HU → relative electron density (piecewise-linear). Radiological depth is ray-traced along divergent rays from a point source at SAD = 100 cm. Dose = TMR-like depth curve (build-up + exponential, per energy) × inverse-square × lateral fluence smoothed by a depth-dependent Gaussian penumbra. IMRT uses 5 mm beamlets and a gradient-based fluence optimizer (PTV under/over-dose, OAR limit and body penalties). No scatter kernel, no heterogeneity lateral transport, no MLC physics, no 3D volume, no couch/skin/contrast handling.
Built-in cases are public, de-identified CT series from The Cancer Imaging Archive (TCIA), cropped and downsampled (224×224) for speed. Thorax: LCTSC (Yang J. et al., CC BY 3.0). Pelvis: Pelvic-Reference-Data (Afua A. et al., CC BY 3.0). Upper abdomen: Spine-Mets-CT-SEG (CC BY 4.0). Data are used for teaching only; please cite TCIA and the original authors. Clark K. et al., J Digit Imaging 2013;26:1045.
Not commissioned, not validated, no MU calculation. Dose is computed on the displayed slice only, so the DVH is a slice-wise (area) DVH. Do not compare numbers with a clinical TPS. DICOM import is client-side only and supports uncompressed CT series; RTSTRUCT/RTDOSE are not read.